Product Recall Reference 2407-0092

FOLO-UP Product Recall: 2407-0092 700HPM High Pressure Calibration Manifold

700HPM High Pressure Calibration Manifold.

Recorded risk Serious Source updated 30 April 2025

Why this product was flagged

Recorded hazard

This product presents a serious risk of injuries as during manual pressurization of the equipment, the plunger nut threads of the High-Pressure Pneumatic Calibration Manifold may yield releasing the plunger assembly and allowing it to be ejected at high velocity. The threads securing the plunger assembly have been found to fail prematurely. This can result in serious injury to the user.

What the notice records

Corrective action

Fluke UK Ltd has stopped shipping the affected manifold and recalled the product from end users. Further information is available on the Fluke website at the following link:https://www.fluke.com/en-gb/support/safetynotices/fluke-700hxx-recall.

Identify the product

Product details

Product
700HPM High Pressure Calibration Manifold
Brand
Fluke
Models
Not provided
Product type
Pressure Equipment / Vessels - High Pressure Calibration Manifold
Seller or shop
Unknown
Country of origin
United States
Measure
Recall of the product from end users
Notified by
Health and Safety Executive

Found this product online?

Flag the listing while the evidence is visible.

A Whistle report is a listing signal submitted for evidence capture and controlled human review. It is not treated as a confirmed listing match until checked.

Get FOLO-UP Whistle