Lena Lighting / UV Cleanlight UVC Sterilon 36w, 72w and 108w
Recalls Reference: 2201-0375
Open recall intelligence view (subscriber access) Intelligence access is currently restricted on public records. To unlock full intelligence access, contact us.
Premium intelligence route retained for subscriber access.
Share this page
- Description
- UV-C lamp advertised as effectively killing microorganisms, used for surface disinfection in a variety of healthcare and public settings.
- Brand
- Lena Lighting / UV Cleanlight
- Shop
- Unknown
- Country of Origin
- Poland
- Risk Level: [Not provided]
- The product has been assessed as unsafe and presents a serious risk because there is a foreseeable risk of exposure to UV-C radiation, which could result in irreversible eye and/or skin injuries. The exposure limit value for this type of radiation is met within 3 seconds of energising the lamp and there are no safeguards in place, such as a motion sensor, to prevent exposure.
- Corrective Action
- We recommend end users stop using this product immediately.
- Product Type
- Electrical equipment - UV - C surface disinfection Lamp
- Document Type
- Product Recall
- Updated